Calibration and verification procedure for the WIKA TR10 RTD sensor

The WIKA TR10 is a Pt100 thin-film platinum assembly, and its calibration decides whether the readings deserve trust. Preparation, a three-step verification flow and tolerance criteria below, with record-keeping essentials.

1. How it works

Before calibrating, three checks: nameplate - Pt100, tolerance class A or B, range against process requirements; appearance - protection tube without dents or cracks, terminals clean and tight, no kinks in the mineral-insulated section; loop - insulation from element to tube above 20 megohms at 500V megger, three-wire balance within limits. Prepare standards: a secondary standard platinum thermometer or precision digital reference (error under one third of the tolerance under test), a stirred bath or dry-block calibrator, an ice-water vessel, and a four-wire measuring instrument, all within their calibration validity - the legal basis of the data.

2. Practice and pitfalls

Step one, ice point: immerse the element and the reference together in a proper ice-water slurry, sensing tips fully covered, read after 10 minutes of stabilisation; a class A unit must read 0±0.15 degrees. Step two, bath comparison: set the bath to 100 degrees or two to three common process points, compare immersed readings, and note deviations; for dry blocks match the insert hole to the element diameter with clearance under 0.5mm, and immersion depth deep enough to defeat stem losses. Step three, data: deviations at each point must sit inside the tolerance band and change smoothly with temperature - a sudden step points at contact or element damage. Disposition: class A units over tolerance but under 0.5 degrees of bias may be downgraded to class B with a note; larger or fast-drifting units are replaced. Records carry five fields: tag number, date, standard reference, values and deviations per point, verdict and next due date. Trend across calibrations: monotonic drift means element aging, erratic jumps mean wiring or insulation. Period: two years generally, annually in hot, wet or vibrating duty, and per the safety-instrument schedule for interlocked points. TR10 thin-film elements drift little in normal service - when a check shows clear drift, suspect mounting and wiring before replacing.

3. TR10 verification flow and tolerances

StepOperationPass criterion
Nameplate and visualType, class, range, tube conditionMatches process, no damage
Insulation test500V megger to the tubeAbove 20 megohms
Ice pointIce slurry, 10 minutes, both immersedClass A: 0±0.15 degrees
Bath comparisonTwo or three process pointsDeviations inside the band, smooth
Data reviewMonotonicity of deviationNo sudden steps
DispositionDowngrade under 0.5 degrees, else replaceVerdict and next date recorded

4. Notes

The value of calibration lives in the trend, not the single pass: one good point proves today only, and the cross-period trend exposes aging. Archive every TR10 curve, schedule replacement on trend alarms, and planned swaps beat emergency trips every time.

5. References

The quantities and component characteristics discussed here follow current national standards and the corresponding IEC documents: Industrial platinum RTD tolerances per GB/T 30121 (IEC 60751); verification per the relevant metrology regulations. Always check the device datasheet and the product design file for ratings and derating.

6. Cautions

⚠️ Three common pitfalls:
① Oversized dry-block holes:A loose thermal bridge falsifies the whole dataset - hole clearance under 0.5mm
② Impure ice point:Ice cubes in tap water shift the zero - use a proper slurry confirmed by the reference
③ Filing one passing certificate and moving on:Gradual drift accumulates unseen - compare every deviation with the previous record

Locate the issue with the approach above, then confirm with measurements; safety-related circuits must follow the standards and manufacturer requirements.

Related reading

Related 3D models: Ready to build it? Download free 3D models & CAD drawings at MOULDL »
MOULDL · Electrical Knowledge Base · free self-study reference for electrical & automation engineers, 3D CAD models at MOULDL