E43-627 · Surface Roughness Mismatch with Fit · A snap-gauge measuring face designed Ra0.05 but measured Ra0.4, causing rapid face wear and reading drift out of tolerance · ★★
The gauge measuring face is too rough at Ra0.4; contact wear is rapid and the reading drifts beyond the calibration tolerance.
[Mechanism] The gauge face contacts the workpiece repeatedly; excessive roughness peaks accelerate wear that changes its dimensions, so the reading drifts with wear and falls out of calibration.
[Calculation] The wear rate at Ra0.4 is several times that at Ra0.05; after several thousand contacts the face wears 0.005, exceeding the snap-gauge calibration tolerance of 0.003.
[Design Omission] The gauge measuring face was left too rough; this design defect causes wear and out-of-tolerance reading.
1) Lap the measuring face to Ra0.025~0.05;
2) Harden or hard-chrome-plate the face for wear resistance;
3) Set Ra per JJG calibration requirements;
4) Measure Ra and reading on the first article;
5) Call out the measuring-face Ra on the drawing;
6) Calibrate periodically;
7) Protect against bumps in use;
8) Monitor wear.
GB/T 131-2006 Geometrical product specifications — Surface texture designation