E43-349 · Thermal Deformation Not Compensated · A granite surface plate 2000mm long on a steel frame, designed isothermal, measured 5℃ temperature change · ★★
The CTE mismatch between slab and frame drives measurement error beyond 0.035mm and biases the readings.
[Mechanism] Granite (α=8e-6) and the steel frame (α=11.5e-6) expand at different rates; the relative displacement introduces error.
[Calculation] With ΔT=5℃, the relative mismatch=(11.5-8)e-6×2000×5=0.035mm, exceeding the measurement tolerance.
[Design Omission] No isothermal soak or material matching — a design defect leading to measurement error.
1) Soak isothermally ≥4h before measuring to eliminate the slab-frame temperature difference;
2) Build a CTE-mismatch compensation model to correct the 0.035mm;
3) Wrap the steel frame with thermal insulation to control temperature;
4) Control temperature within ±0.5℃ to reduce ΔT;
5) Use low-expansion frame material matched to granite;
6) Calibrate readings with a reference standard;
7) Use an Invar reference scale to reduce the mismatch;
8) Re-measure the first article to verify error ≤0.01mm.
GB/T 16857.2-2017 Coordinate measuring machines (CMM) — Acceptance and re-verification tests