E45-304 · Calibration and Adjustment · Temperature Transmitter · ★
The built-in self-calibration function was left off; the 0.3%FS daily zero drift was not cancelled automatically
[Mechanism] The transmitter has a built-in self-calibration circuit for periodic automatic zeroing; leaving it off lets the daily zero drift accumulate and need manual intervention to correct.
[Calculation] The daily drift is 0.05%FS; with self-calibration off it accumulates to 1.5%FS in a month, over the limit, while with it on the drift is cancelled daily.
[Design Omission] Self-calibration was not enabled and no interval was set.
① Enable self-calibration: switch on the built-in self-calibration → cancel the daily drift
② Set the interval: set the self-calibration interval to daily → timely correction
③ Calibration monitoring: monitor self-calibration execution → prevent omissions
④ Verification: do a periodic manual recalibration → confirm consistency
⑤ Out-of-tolerance alarm: warn when the zero deviation >0.3% → early warning
⑥ Records: keep the self-calibration log on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the self-calibration requirement into the procedure → control at the source
GB/T 17614.1-2015 Industrial process control systems - Transmitters - Part 1: Methods of evaluating the performance