E45-470 · Vibration and Mechanical Stress · Temperature Sensor · ★★★
Vibration fatigues and fractures the thermocouple protection tube; the thermocouple is exposed and damaged, and temperature measurement fails
[Mechanism] Under vibration the protection tube undergoes alternating bending fatigue; cracks propagate from the stress concentration at the root until it fractures.
[Calculation] The protection tube stress exceeds the fatigue limit and it fractures after thousands of hours, exposing and damaging the thermocouple wires.
[Design Deficiency] The protection tube lacked stiffness and had no vibration damping.
① Increase stiffness: thicken the protection tube to raise strength → resist fracture
② Damped mounting: add elastic supports to the thermocouple → reduce stress
③ Wear-resistant material: choose a fatigue-resistant material → extend service life
④ Limit constraint: add constraints to limit sway → prevent fracture
⑤ Regular inspection: check for tube cracks → early correction
⑥ Tube-fracture monitoring: monitor temperature anomalies → early detection
⑦ Fracture verification: confirm with a vibration endurance test → data confirmation
⑧ Selection standardization: write the protection tube strength into the spec → control at the source
GB/T 16839.1-1997 Thermocouples - Part 1: Reference tables