E41-904 · Seal Component Selection · Pharmaceutical Filling Valve · ★★★★
One pharmaceutical valve used plain EPDM; after steam sterilization, extractables exceeded limits and the batch failed.
First, plain EPDM contains extractables; 121°C steam sterilization promotes migration beyond pharmacopoeia limits. Second, extractables contaminate the drug solution, failing test items. Third, whole batch scrapped + compliance risk. Root cause: non-pharma-grade.
① Pharma/medical use must use USP Class VI / medical-grade silicone or EPDM, steam-sterilizable
② Material must have low extractables and biocompatibility certificate
③ Do not use plain industrial rubber in pharma
④ Periodically test extractables
⑤ Keep a material certificate ledger
⑥ Set sterilization cycle by material
⑦ If failing, check extractables immediately