E43-492 · Dimensional-Chain Accumulation · CMM X-direction designed ±0.003, measured ±0.015 after 3 axes · ★★★
Multi-axis positioning accumulation made X-direction measurement 0.015mm out of tolerance, biasing readings.
[Mechanism] Multi-axis error dimensional chain accumulated, so the closed loop (X-direction) went out of tolerance.
[Calculation] T0=√(3×0.009²)=0.015, 5× the 0.003 allowance.
[Design Omission] Per-axis accuracy not allocated; design defect → X-direction out of tolerance.
1) Allocate each axis accuracy per T0/√3;
2) Linear-scale compensation;
3) Calibrate each axis;
4) Statistical check;
5) Selective matching by measurement;
6) Leave adjustment allowance;
7) Standard artifact calibration;
8) First-article X-direction measured ≤0.003.
GB/T 16857.2-2017 Coordinate measuring machines (CMM) — Acceptance and re-verification tests