Excessive CT Gantry Bearing Bore Roundness Causes Image Reconstruction Artefacts

E43-294 · Rolling Bearing Fit Errors · Medical CT Gantry Bearing 7016C ∅80, housing bore roundness designed ≤0.01, actual 0.04 · ★★

Problem Symptom

A bore roundness of 0.04mm (allowable 0.01) made the gantry run out 0.04 and the reconstructed images showed artefacts

Root Cause

[Mechanism] The out-of-round bore deformed the bearing and the gantry ran out radially.
[Calculation] A roundness of 0.04 gave a gantry runout of 0.04, producing reconstruction artefacts.
[Design Omission] The bore roundness was not controlled; this design flaw caused the artefacts.

Correct Solution

1) Control the bore roundness ≤0.01 to preserve concentricity;
2) Finish bore and lap;
3) Select P4 grade for higher precision;
4) Hot assemble to prevent damage;
5) Inspect the radial runout ≤0.005 after installation;
6) Dynamically balance the gantry;
7) Trial the scanning;
8) Bring the roundness onto the drawing.

Checklist

References

GB/T 1182-2018 Geometrical Tolerances

Related 3D models: Ready to build it? Download free 3D models & CAD drawings at MOULDL »
MOULDL · Mechanical Engineering Knowledge Base · free self-study reference for machine designers, 3D CAD models at MOULDL