E43-919 · Seal and Clearance Fit Error · An instrument fitting metal seal face designed roughness Ra0.4 measured Ra3.2, so microscopic peak-valley paths on the metal contact leak gas microscopically · ★★
A gas-analyzer instrument metal-to-metal seal fitting had a sealing surface roughness measured at Ra3.2, above the designed Ra0.4; microscopic peaks and valleys on the metal-to-metal face form connected leak paths, gas micro-leaks, and the leak test fails.
[Mechanism] A metal-to-metal seal relies on high face conformity to block leakage; excessive roughness connects microscopic peaks and valleys into capillary leak paths, and gas micro-leaks along them.
[Calculation] Micro-leak rate ∝ peak-valley cross-section; raising Ra from 0.4 to 3.2 (8×) enlarges the peak-valley paths, and the micro-leak rate rises sharply past the leak-test limit.
[Design Omission] The seal face was left as-machined without fine grinding — a design defect leading to over-rough surface and micro leak.
1) Grind metal seal faces to Ra0.2~0.4 to cut peak-valley paths;
2) On first articles, check roughness with a profilometer, holding the upper limit;
3) Control flatness together with roughness;
4) Add a soft-metal gasket to fill peaks/valleys if needed;
5) Size micro-leak ∝ roughness, quantitatively;
6) On first articles, pass a helium leak test, to verify;
7) Call out seal-face Ra and flatness on the drawing;
8) If the seal face is over-rough, re-grind to conform.
GB/T 1031-2009 Surface roughness — Parameters and values