Knowledge Base › Tolerance and Fit Error Case › CMM Probe Thermal Bending - Measurement Error Out

CMM Probe Thermal Bending - Measurement Error Out

E43-320 · Thermal Deformation Uncompensated · A CMM probe 300mm long in steel, designed isothermal, measured 4℃ change · ★★

Problem Description

Probe temperature change of 4℃ elongated 0.014mm, making measurement error out of tolerance and readings unstable.

Root Cause

[Mechanism] The probe axially expands when heated and shifts the stylus contact point, directly introducing length measurement error.
[Calculation] α=11.5e-6, Δl=11.5e-6×300×4=0.0138mm≈0.014mm, exceeding the 0.01mm measurement tolerance.
[Design Omission] No isothermal balancing or temperature compensation — a design defect leading to measurement error out of tolerance.

Corrective Solution

1) Isothermal-balance ≥2h before measuring so the probe and workpiece match temperature;
2) Build a probe-temperature compensation to correct the 0.014mm;
3) Hold temperature ±0.5℃ to cut ΔT;
4) Select low-expansion probe material α≈3e-6 to cut Δl;
5) Monitor probe temperature in real time;
6) Calibrate readings with gauge blocks;
7) Periodically re-check;
8) On first articles, re-measure and verify error ≤0.005mm.

Checklist

Reference

GB/T 16857.2-2017 Coordinate measuring machines (CMM) — Acceptance and re-verification tests