Ruling-Machine Scale/Base CTE Mismatch - Ruling Out

E43-331 · Thermal Deformation Uncompensated · A ruling-machine steel scale 1000mm long mounted on an aluminum base, designed isothermal, measured 3℃ change · ★★

Problem Description

The CTE mismatch between the scale and base made ruling error go 0.02mm out of tolerance and readings off.

Root Cause

[Mechanism] The steel scale α=11.5e-6 and aluminum base α=23e-6 expand out of step; the relative displacement introduces locating error.
[Calculation] ΔT=3℃, relative elongation difference=(23−11.5)e-6×1000×3=0.0345mm≈0.035mm out of tolerance.
[Design Omission] No same-material or isothermal compensation — a design defect leading to ruling out of tolerance.

Corrective Solution

1) Isothermal-balance ≥3h before measuring to remove scale-base temperature difference;
2) Build a CTE-mismatch compensation model to correct the 0.035mm;
3) Change the scale base to steel or low-expansion material to match;
4) Control temperature within ±0.5℃ to reduce ΔT;
5) Monitor scale temperature in real time;
6) Calibrate readings with a reference standard;
7) Select invar scale to cut the difference;
8) Re-measure the first article to verify error ≤0.01mm.