E44-1115 · Miscellaneous · Rotameter · ★★
Backward install, reading, offset
[Mechanism] The rotameter is installed backward (taper tube's small end down), so the float cannot rise and the reading is false.
[Calculation] Correct mounting has the small end up and large end down; reversed, the float sits on the bottom and reads 0 (actual 30m³/h), a 100% error.
[Design Omission] No flow-direction check, no marking, no through-flow test, and no verticality control.
1) Check direction and install correctly (small end up), since reversal causes error and correct mounting reads true, restoring accuracy;
2) Add markings (flow-direction arrows), since ambiguity leads to reversal and markings remove doubt, preventing wrong installation;
3) Run a through-flow test (float rises with flow), since without proof direction is unknown and a measured test confirms it, setting installation on data;
4) Hold verticality (≤2mm/m), since tilt causes error and a true vertical reads correctly, giving a true reading;
5) Detect reversed mounting (shows up when installed backward), since reversal causes gradual error and a check reveals it, monitoring installation;
6) Fit reading monitoring (alarm on offset), since reversal first shows as a zero reading for early warning, monitoring deviation;
7) Set a calibration interval (every six months), since long service drifts and scheduled calibration prevents it, guarding against reversal in service;
8) Write direction check and marking into the flowmeter card, since reversal stems from no direction check and no marking, closing the loop on the card.
GB/T 2624.1-2006 Flow measurement