E44-491 · Filter Blockage and Fouling · Dialysis-Water Filter 0.1μm · ★★★
Endotoxin 0.5EU/ml > 0.25, exceeds the national standard, dialysis safety risk
[Mechanism] A biofilm on the membrane releases endotoxins.
[Calculation] Membrane aging lowers rejection, endotoxins break through, outlet 0.5 exceeds 0.25.
[Design Omission] No scheduled membrane replacement and no thermal sanitization; the limit is exceeded.
1) Replace the membrane on schedule to preserve rejection, because the original aging, lowered rejection and 0.5 exceeding 0.25 caused the overrun, so membrane replacement controls endotoxin;
2) Periodically hot-water sanitize to kill membrane bacteria, because bacteria release endotoxins, so thermal sanitization is used;
3) Online endotoxin monitoring of the outlet, because an overrun is hazardous, so monitoring is provided;
4) Add upstream ultrafiltration to reduce load, because the load is high, so upstream filtration is used;
5) Design piping dead-leg free to prevent bacteria, because dead legs breed bacteria, so dead-leg-free design is used;
6) Flush regularly to prevent buildup, because buildup grows bacteria, so scheduled flushing is used;
7) Keep batch records for traceability, because medical use requires traceability, so traceability is maintained;
8) Write membrane replacement and sanitization into the dialysis specification, because the lack of replacement was the root cause of endotoxin, so the dialysis spec governs.