E45-096 · Signal Interference and EMC · Differential Pressure Transmitter · ★★
Under common-mode mains-frequency interference the reading has ±0.4%FS superimposed on it, with an obvious 50Hz component
[Mechanism] The front-end amplifier's common-mode rejection ratio is insufficient, so mains-frequency common-mode interference is not rejected and turns into differential mode added to the reading.
[Calculation] With common-mode voltage Vcm=5V and CMRR=60dB, the residual differential Vd=Vcm/10^3=5mV corresponds to ±0.4%FS; CMRR needs to be ≥100dB to press it to ≤0.05%.
[Design Omission] A low-CMRR op-amp was used, with no balancing and no shielding.
① High CMRR: change the front end to an instrumentation amplifier with CMRR≥100dB → common-mode-to-differential conversion down 100×
② Balance: make the two input impedances strictly symmetric (difference ≤0.1%) → suppresses common mode
③ Shielded twisted pair: change the signal cable to shielded twisted pair → less common-mode pickup
④ Mains notch: add a 50Hz notch filter (≥40dB) → filters out the mains frequency
⑤ Isolation: signal isolation (≥1kV) → cuts common mode
⑥ Single-end grounding: ground the shield at one point → avoids ground loops
⑦ Verify by measurement: readings in a mains-frequency field (differential ≤0.05%FS) → confirmed
⑧ Institutionalize: write the CMRR requirement into the selection sheet → gate at the source
GB/T 17626.6-2017 Immunity to conducted disturbances induced by radio-frequency fields