E45-097 · Signal Interference and EMC · Strain-Gauge Force Sensor · ★★
After the ambient fan starts or stops, the force reading drifts ±0.2%FS (normally ≤0.02%FS)
[Mechanism] Unshielded, untwisted strain-bridge leads pick up mains-frequency and RF interference, which at high gain shows up as apparent thermal drift.
[Calculation] With an untwisted two-wire loop area A≈10cm² and mains-frequency B≈0.3mT, the induced e≈2πfBA≈56μV, amplified by a gain of 1000 into a ±0.2%FS drift.
[Design Omission] The strain leads were parallel wires, unshielded and not kept away from interference sources.
① Twisted shielded pair: change the strain leads to twisted shielded pair with the shield grounded at one end → loop area approaches zero, induction down 90%
② Move away from the source: keep at least 300mm from the VFD/fan → reduces field strength
③ High CMRR: use bridge amplification with CMRR≥100dB → suppresses common mode
④ Low-pass filtering: add a low-pass filter (10Hz cutoff) → filters the high frequency
⑤ Shield can: shield the junction box with metal → blocks radiation
⑥ Temperature compensation: add temperature compensation to the bridge → separates out the true thermal drift
⑦ Verify by measurement: readings with the fan starting/stopping (drift ≤0.02%FS) → confirmed
⑧ Institutionalize: write the strain-cable type into the standard → gate at the source
GB/T 17626.6-2017 Immunity to conducted disturbances induced by radio-frequency fields