E45-192 · Grounding and Shielding · Laboratory Measuring Instrument · ★★
After the measuring instrument was powered from two different grounds, the reading jumps, indicating an obvious ground circulating current
[Mechanism] The instrument's power ground and signal ground form a loop through the PE of two socket outlets, and the ground potential difference drives a circulating current that drops voltage in the signal circuit; an isolation transformer cuts the primary-to-secondary ground connection, and the floating secondary leaves the circulating current no path.
[Calculation] Ground potential difference ΔU=3V across a loop resistance R=0.5Ω → circulating current I=6A, which drops an appreciable voltage across the signal source impedance and causes the jump; with an isolation transformer I≈0.
[Design Omission] In a high-interference situation no isolated supply or isolation amplifier was used.
① Isolation transformer: add an isolation transformer to the instrument supply (2500V isolation) → cuts the primary-to-secondary ground loop
② Floating secondary: leave the secondary unconnected to PE, with the signal ground referenced at a single point → no path for circulating current
③ Isolation amplifier: add an isolation amplifier to the signal channel → decouples the grounds
④ Single-point grounding: bond the secondary signal ground to the local PE at one point → establishes a reference without forming a loop
⑤ Equipotential bonding: still bond the two grounds to lower ΔU → a second safeguard
⑥ Loop-current test: after adding isolation, measure the circulating current ≈0 → verify
⑦ Jump ratio: reading jumps <1LSB → verify
⑧ Write the isolated supply into the high-precision measurement specification → gate at the source
GB/T 12179-2008 Electrical voltage dividers