E45-220 · Grounding and Shielding · Strain Gauge Measurement Leads · ★
The strain measurement jumped when the inverter started and stopped because the leads were ordinary parallel wires with no shielding
[Mechanism] A strain gauge outputs only millivolts, and ordinary parallel wires offer neither twisted-pair cancellation nor shielding; the ambient electromagnetic field couples straight into the pair as differential-mode noise, and the transient field from inverter switching is especially strong.
[Calculation] With an interference field E=20V/m, a 2m parallel pair and a loop area of 0.002m², the induced e≈several millivolts, far above the strain signal (2mV/με).
[Design Omission] Ordinary parallel wire was wrongly used for the low-level strain measurement.
① Shielded twisted pair: change the strain leads to shielded twisted pair → reduces coupling
② Single-end shield: bond the shield to PE at the receiver end only → establishes a reference
③ Away from interference: keep the wiring >30cm from the inverter → lower field
④ Metal conduit: run the low-level signal in a grounded metal conduit → second shield
⑤ Low-pass filter: add an RC low-pass at the front end (cut-off 10Hz) → suppress transients
⑥ Noise test: jump <0.5με when the inverter starts and stops → confirmed
⑦ Comparison: noise falls >10 times after rewiring → confirmed
⑧ Write the low-level signal cable type into the measurement specification → control at the source
GB/T 13992-2010 Metallic bonded resistance strain gauges