E45-243 · Wiring and Terminals · Outdoor Temperature Transmitter · ★
The junction box gasket aged and let water in; the terminals corroded and the signal jumped
[Mechanism] The junction box gasket aged or the cover was not tightened; rainwater seeped in along the leads, the terminals and wires became damp and rusted, the contact resistance grew, and leakage added noise to the signal.
[Calculation] The water ingress drops the insulation between terminals from >100MΩ to 1MΩ; the leakage current produces ±0.5% mains-frequency interference in the signal circuit and the reading jitters.
[Design Omission] No waterproof gland, a gasket material that does not weather, and no downward cable entry.
① Waterproof gland: seal the lead entry with a cable gland → no ingress
② Downward entry: bring cables out at the bottom so drips cannot collect → drain
③ Weatherproof gasket: switch to a fluoroelastomer gasket → resist aging
④ Seal inspection: check for no water traces → early prevention
⑤ Insulation measurement: measure >10MΩ → confirmation
⑥ Desiccant: put desiccant in the box → keep out damp
⑦ Corrosion check: inspect the terminals periodically → replace early
⑧ Write the sealing requirement into installation → control at the source
GB 4208-2017 Degrees of protection provided by enclosures (IP code)