E45-287 · Calibration · Temperature Transmitter · ★★
A class-1.0 reference gauge was used to calibrate a class-0.2 sensor; the post-calibration error reached 0.8%FS
[Mechanism] The reference standard's accuracy is lower than that of the unit being calibrated, so the calibration uncertainty exceeds the tolerance and the sensor's deviation cannot be corrected properly.
[Calculation] The class-1.0 gauge has an uncertainty of 1.0%FS while the class-0.2 unit has a tolerance of 0.2%FS; the ratio is 5 times, exceeding the required 3 times, leaving a 0.8% error after calibration.
[Design Omission] The rule that the standard must be 3 times better than the unit was not followed and the class was not checked.
① Class matching: the standard ≥3 times better than the unit → true value
② Validity check: check that the certificate is current → accurate reference
③ Self-check: self-check the standard periodically → keep it accurate
④ Recheck: re-calibrate with another standard → confirm agreement
⑤ Out-of-tolerance alarm: forbid calibration when the standard is out of tolerance → prevent error
⑥ Certificates: keep the certificate numbers → traceability
⑦ Management: keep a standards log → manage the intervals
⑧ Write the class requirement into the rules → control at the source
JJF 1001-2011 General Terms in Metrology and Their Definitions