Knowledge Base › Electrical Sensor Error Case › Wrong Ice Point in Temperature Calibration

Wrong Ice Point in Temperature Calibration

E45-299 · Calibration and Adjustment · PT100 Platinum Resistance · ★

Problem Description

Using impure ice water to calibrate 0℃, the ice point is off by 0.3℃ and the whole range shifts

Root Cause

[Mechanism] The ice-water mixture used for ice-point calibration is impure or not stirred enough, so the actual temperature is not 0℃ (about 0.3℃); as the zero reference it translates the whole range.
[Calculation] An ice-point error of 0.3℃ shifts the zero reference by 0.3℃, translating the whole reading by 0.3℃.
[Design Omission] Pure crushed ice was not used and stirred enough, and the ice point was not measured.

Corrective Solution

① Pure-ice calibration: use pure crushed ice with stirring → true 0℃
② Ice-point measurement: measure and confirm the ice point first → accurate reference
③ Isothermal calibration: keep the calibration environment stable → no drift
④ Verification: re-measure the zero after calibration → confirm accuracy
⑤ Out-of-tolerance alarm: warn when the deviation >0.1℃ → early warning
⑥ Records: keep the ice-point state on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the ice-point requirement into the procedure → control at the source

Checklist

Reference

JJG 229-2010 Verification Regulation of Industrial Platinum Resistance Thermometers

Related model download:Understood the principle and ready to try?Go to MOULDL to download related 3D models and CAD drawings for free »