E45-298 · Calibration and Adjustment · Temperature Transmitter · ★
The medium was not drained before calibration; the residual pressure shifts the zero by 2%FS
[Mechanism] The measuring chamber was not cleared and evacuated before calibration; residual medium pressure or temperature shifts the initial reference away from zero, so the zero calibrated this way is wrong.
[Calculation] A residual pressure of 2kPa corresponds to a 2%FS zero deviation; the whole range translates by 2%.
[Design Omission] No pre-calibration clearing was done and the zero was not confirmed.
① Pre-clearing: drain and clear before calibration → correct reference
② Zero confirmation: confirm the zero is stable → then calibrate
③ Procedure: write the clearing requirement → prevent omissions
④ Verification: re-zero after calibration → confirm accuracy
⑤ Out-of-tolerance alarm: alarm on an unstable zero → prevent a bad calibration
⑥ Records: keep the cleared state on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the clearing requirement into the procedure → control at the source
GB/T 17614.1-2015 Industrial process control systems - Transmitters - Part 1: Methods of evaluating the performance