E45-301 · Calibration and Adjustment · Pressure Transmitter · ★
Dust entering the calibration port contaminates the diaphragm; the zero drifts by 1%FS
[Mechanism] The port is not sealed during calibration, so dust or oil enters and contaminates the diaphragm and the pressure-sensing cavity, changing the zero characteristic and shifting the calibration reference.
[Calculation] The contamination loads the diaphragm unevenly and the zero shifts by 1%FS; the whole range translates by 1%.
[Design Omission] No port protection was provided and nothing was cleaned.
① Seal the port: seal the port before calibration to keep out contamination → clean diaphragm
② Cleaning: clean the pressure-sensing cavity before calibration → no contamination
③ Dust protection: keep the calibration bench dust-free → no dust entry
④ Verification: re-zero after calibration → confirm accuracy
⑤ Out-of-tolerance alarm: warn when the zero deviation >0.5% → early warning
⑥ Records: keep the cleanliness state on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the protection requirement into the procedure → control at the source
GB/T 17614.1-2015 Industrial process control systems - Transmitters - Part 1: Methods of evaluating the performance