Knowledge Base › Electrical Sensor Error Case › Inaccurate Vibration Calibration Table

Inaccurate Vibration Calibration Table

E45-302 · Calibration and Adjustment · Accelerometer · ★★

Problem Description

The calibration table's own frequency response is uneven; the calibrated sensitivity is off by 8%

Root Cause

[Mechanism] At the calibration frequency the vibration calibration table has an inaccurate amplitude or large transverse vibration; as the excitation reference this error makes the sensitivity coefficient of the accelerometer under test wrong.
[Calculation] The table amplitude is off by 8% at that frequency, so the calibrated sensitivity is off by 8% and the whole band is distorted.
[Design Omission] The calibration table was not checked and the transverse vibration was not controlled.

Corrective Solution

① Table check: check the calibration table's frequency response before calibration → true reference
② Transverse control: keep the transverse vibration <5% → single-axis accuracy
③ Grade matching: the table grade is three times that of the unit under test → true value
④ Verification: recalibrate on another table → confirm consistency
⑤ Out-of-tolerance alarm: warn when the sensitivity deviation >5% → early warning
⑥ Records: keep the table condition on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the table-check requirement into the system → control at the source

Checklist

Reference

JJG 233-2008 Verification Regulation of Piezoelectric Accelerometers