E45-306 · Calibration and Adjustment · Pressure Transmitter · ★
The temperature-drift compensation coefficient has the wrong sign; as the temperature rises the reading goes up by 2%FS instead
[Mechanism] The temperature compensation coefficient was written with the wrong sign; on temperature rise the compensation adds in the reverse direction, doubling the drift that should have been cancelled.
[Calculation] The true temperature drift is +0.02%/℃ but the compensation is wrongly -0.02%; a 20℃ rise compounds it to a 2%FS shift in the reverse direction.
[Design Omission] The sign of the compensation coefficient was not checked and no temperature-change verification was done.
① Sign check: check the sign of the temperature-compensation coefficient → correct direction
② Temperature-change verification: test both heating and cooling → verify compensation
③ Stability criterion: after compensation the temperature-change deviation <0.1% → accurate
④ Verification: compare at different temperature points → accurate throughout
⑤ Out-of-tolerance alarm: warn on reverse-direction deviation → early warning
⑥ Recalibration: recalibrate when wrong → correct the bias
⑦ Records: keep the coefficient on file → traceability
⑧ Write the compensation requirement into the procedure → control at the source
GB/T 17614.1-2015 Industrial process control systems - Transmitters - Part 1: Methods of evaluating the performance