E45-309 · Calibration and Adjustment · Load Cell · ★
Long-term stability was not verified; the 1%FS zero drift was only found out of tolerance after half a year
[Mechanism] Calibration verified only the instantaneous accuracy and did not assess long-term stability; creep and aging only appear in later operation, so the drift is not exposed in advance.
[Calculation] No 48h stability test was done; the creep accumulated to 1%FS over half a year, exceeding the 0.5% limit, before it was found.
[Design Omission] No long-term stability test was done and no tracking was kept.
① Stability test: run a 48h stability test → expose the drift in advance
② Creep check: keep creep <0.1%/30min → stable
③ Tracking: log the drift over the long term → predict replacement
④ Periodic calibration: shorten the interval in the first year → prevent drift
⑤ Out-of-tolerance alarm: warn when the drift >0.5% → early warning
⑥ Aging management: replace on the rated life → no accumulation
⑦ Records: keep the stability-test data on file → traceability
⑧ Write the stability-test requirement into the procedure → control at the source
JJG 669-2003 Verification Regulation of Load Cells