Knowledge Base › Electrical Sensor Error Case › Long-Term Stability Not Verified

Long-Term Stability Not Verified

E45-309 · Calibration and Adjustment · Load Cell · ★

Problem Description

Long-term stability was not verified; the 1%FS zero drift was only found out of tolerance after half a year

Root Cause

[Mechanism] Calibration verified only the instantaneous accuracy and did not assess long-term stability; creep and aging only appear in later operation, so the drift is not exposed in advance.
[Calculation] No 48h stability test was done; the creep accumulated to 1%FS over half a year, exceeding the 0.5% limit, before it was found.
[Design Omission] No long-term stability test was done and no tracking was kept.

Corrective Solution

① Stability test: run a 48h stability test → expose the drift in advance
② Creep check: keep creep <0.1%/30min → stable
③ Tracking: log the drift over the long term → predict replacement
④ Periodic calibration: shorten the interval in the first year → prevent drift
⑤ Out-of-tolerance alarm: warn when the drift >0.5% → early warning
⑥ Aging management: replace on the rated life → no accumulation
⑦ Records: keep the stability-test data on file → traceability
⑧ Write the stability-test requirement into the procedure → control at the source

Checklist

Reference

JJG 669-2003 Verification Regulation of Load Cells

Related model download:Understood the principle and ready to try?Go to MOULDL to download related 3D models and CAD drawings for free »
MOULDL · Mechanical Engineering Knowledge Base · free self-study reference for machine designers, 3D CAD models at MOULDL