Knowledge Base › Electrical Sensor Error Case › Calibration Load Mismatch

Calibration Load Mismatch

E45-312 · Calibration and Adjustment · Load Cell · ★

Problem Description

A large scale is calibrated with a small load; the full-scale linearity is off by 3%FS

Root Cause

[Mechanism] The standard load used for calibration is far below the range; coefficients calibrated over a small span extrapolate inaccurately to the heavy-load segment, and the full-scale end-point linearity error is large.
[Calculation] Calibrating with only a 20%FS load and extrapolating to 100%FS gives a 3%FS linearity error.
[Design Omission] Calibration did not use a load close to full scale, and the full range was not verified.

Corrective Solution

① Full-load calibration: calibrate with a load ≥80%FS → accurate end points
② Full-range verification: check at the full-scale point → verify linearity
③ Staged calibration: stage large and small loads → each segment accurate
④ Out-of-tolerance alarm: warn when the linearity error >1% → early warning
⑤ Verification: compare against an actual weighing → verify
⑥ Records: keep the loads on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the load requirement into the procedure → control at the source

Checklist

Reference

JJG 669-2003 Verification Regulation of Load Cells