Calibration Without Humidity Compensation

E45-313 · Calibration and Adjustment · Humidity Sensor · ★

Problem Description

The calibration made no humidity compensation; at high field humidity the reading is off by 4%RH

Root Cause

[Mechanism] A humidity-sensing element is nonlinear and varies with humidity; calibrating at only a single humidity point builds no full-range compensation, so the high-humidity segment departs from the calibration curve.
[Calculation] Single-point calibration misses the high-humidity segment; at 80%RH in the field the reading is off by 4%RH, exceeding the 2% limit.
[Design Omission] No multi-point humidity compensation was done and no compensation table was built.

Corrective Solution

① Multi-point humidity: calibrate at different humidity points → accurate across the range
② Humidity compensation table: build a humidity compensation table → remove the segment deviation
③ Verification: compare under high field humidity → confirm consistency
④ Out-of-tolerance alarm: warn when the deviation >2% → early warning
⑤ Periodic calibration: calibrate every quarter → prevent drift
⑥ Records: keep the compensation on file → traceability
⑦ Recalibration: recalibrate on any anomaly → correct the bias
⑧ Write the humidity-compensation requirement into the procedure → control at the source

Checklist

Reference

JJG 875-2019 Verification Regulation of Digital Thermo-Hygrometers