E45-314 · Calibration and Adjustment · Magnetic Sensor · ★
The magnetic field in the calibration coil is non-uniform; sensitivity calibration scatters by 3%
[Mechanism] The uniform-field region at the center of the electromagnetic calibration coil is small; placing the sensor at the edge makes the actual field lower than nominal, so the sensitivity coefficient is calibrated against the wrong field strength.
[Calculation] The field at the edge is 3% lower, biasing the sensitivity by 3% and widening the batch scatter.
[Design Omission] The field's uniform region was not verified and the sensor was not positioned at the center.
① Uniform-region calibration: place at the center of the coil's uniform region → true field
② Field check: verify the magnetic field strength before calibration → accurate reference
③ Positioning: fix the sensor position → good repeatability
④ Verification: recalibrate at a different position → confirm consistency
⑤ Out-of-tolerance alarm: warn when the sensitivity deviation >2% → early warning
⑥ Records: keep the field strength on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the field-uniformity requirement into the procedure → control at the source
GB/T 13823.1-2005 Methods for the calibration of vibration and shock sensors