Knowledge Base › Electrical Sensor Error Case › Wrong Reference Face for Angle Calibration

Wrong Reference Face for Angle Calibration

E45-315 · Calibration and Adjustment · Tilt Sensor · ★

Problem Description

The calibration reference face is not level; the zero shifts by 1° and the angle is off across the full range

Root Cause

[Mechanism] Angle calibration treats a non-level reference face as 0°; a tilted reference face adds a fixed bias to every angle measurement.
[Calculation] A 1° tilt of the reference face shifts the full-range angle reading by 1°, and the positioning error grows with the angle.
[Design Omission] No level was used to verify the reference and it was not leveled.

Corrective Solution

① Reference leveling: verify the reference face with a level before calibration → true 0°
② Leveling: level to <0.05° → accurate reference
③ Verification: compare forward and reverse placements after calibration → confirm symmetry
④ Out-of-tolerance alarm: warn when the zero offset >0.2° → early warning
⑤ Records: keep the reference state on file → traceability
⑥ Periodic calibration: calibrate every quarter → prevent drift
⑦ Recalibration: recalibrate on any anomaly → correct the bias
⑧ Write the reference requirement into the procedure → control at the source

Checklist

Reference

GB/T 13992-2010 Metallic bonded resistance strain gauges

Related model download:Understood the principle and ready to try?Go to MOULDL to download related 3D models and CAD drawings for free »