E45-323 · Calibration and Adjustment · Temperature Transmitter · ★
The interpolation nodes are out of order; the table lookup returns wrong values, off by 2%FS
[Mechanism] The interpolation nodes stored from multi-point calibration were not sorted by the independent variable; the table lookup for linear interpolation picks the wrong interval between out-of-order nodes and returns wrong outputs.
[Calculation] The out-of-order nodes make the interpolation land in the wrong segment, and the reading jumps by 2%FS.
[Design Omission] The nodes were not sorted and monotonicity was not checked.
① Sorted storage: sort the nodes by the independent variable → correct table lookup
② Monotonicity check: verify the nodes are monotonic → no jumps
③ Verification: compare at the interpolation points → confirm accuracy
④ Out-of-tolerance alarm: warn on jumps → early warning
⑤ Recalibration: recalibrate when wrong → correct the bias
⑥ Records: keep the nodes on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the node requirement into the procedure → control at the source
GB/T 17614.1-2015 Industrial process control systems - Transmitters - Part 1: Methods of evaluating the performance