Knowledge Base › Electrical Sensor Error Case › Calibration Curve Extrapolation

Calibration Curve Extrapolation

E45-322 · Calibration and Adjustment · Differential Pressure Transmitter · ★

Problem Description

The operating point exceeds the calibration upper limit; extrapolating the curve gives a 4%FS deviation

Root Cause

[Mechanism] The actual duty exceeds the range of the calibration curve; extrapolating by the end-segment slope amplifies the end-point deviation, and the over-range segment is inaccurate.
[Calculation] The operating point is 120%FS against a calibrated 100%FS; extrapolation gives a 4%FS deviation, over the limit.
[Design Omission] The calibration range was not extended and the operating range was not limited.

Corrective Solution

① Extended-range calibration: calibrate to cover the operating upper limit → no overrun
② Range limiting: limit operation to ≤100%FS → no deviation
③ Over-range alarm: warn when over range → early warning
④ Verification: recalibrate the end points → confirm accuracy
⑤ Choose a larger range: select a larger span → leave margin
⑥ Records: keep the range on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the range requirement into the procedure → control at the source

Checklist

Reference