Knowledge Base › Electrical Sensor Error Case › Excessive Automatic Zero Tracking

Excessive Automatic Zero Tracking

E45-321 · Calibration and Adjustment · Load Cell · ★

Problem Description

The zero tracking is too fast and swallows the true no-load zero; the empty scale shows -0.3%FS

Root Cause

[Mechanism] The automatic zero-tracking threshold and time constant are too aggressive; slow real load changes are mistaken for drift and cancelled, shifting the empty-scale reference.
[Calculation] The tracking rate is so fast that it swallows 0.05%FS per minute; with the scale empty and at rest the display shows a false zero of -0.3%FS.
[Design Omission] The tracking parameters were too aggressive and no lower limit was set.

Corrective Solution

① Parameter softening: set a moderate tracking rate → do not swallow real loads
② Lower-limit setting: set a tracking lower threshold → preserve the true zero
③ Zero check: verify the empty scale reads 0 → correct reference
④ Verification: retest under load → confirm accuracy
⑤ Out-of-tolerance alarm: warn when the zero offset >0.2% → early warning
⑥ Records: keep the parameters on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the tracking parameters into the procedure → control at the source

Checklist

Reference

JJG 669-2003 Verification Regulation of Load Cells

Related model download:Understood the principle and ready to try?Go to MOULDL to download related 3D models and CAD drawings for free »