E45-320 · Calibration and Adjustment · Pressure Transmitter · ★★
The calibration coefficients were not saved to EEPROM; they are lost on power-down, reverting to factory defaults and causing a 2%FS deviation
[Mechanism] The calibration coefficients were not written to non-volatile memory; after power-down the factory defaults are restored, and field measurements are biased by the wrong coefficients.
[Calculation] Not saving to EEPROM makes the coefficients revert to defaults on power-down, differing by 2%FS; the entire range reads 2% off.
[Design Omission] Nothing was written to non-volatile memory, and the save was not verified.
① Save to EEPROM: write to non-volatile memory after calibration → no loss on power-down
② Save verification: read back after writing → confirm it was saved
③ Power-down test: simulate a power-down and retest → verify retention
④ Out-of-tolerance alarm: warn on abnormal coefficients → early warning
⑤ Records: keep the coefficient versions on file → traceability
⑥ Periodic calibration: calibrate every quarter → prevent drift
⑦ Recalibration: recalibrate after any loss → correct the bias
⑧ Write the storage requirement into the procedure → control at the source
GB/T 17614.1-2015 Industrial process control systems - Transmitters - Part 1: Methods of evaluating the performance