Knowledge Base › Electrical Sensor Error Case › Atmospheric Pressure Not Compensated

Atmospheric Pressure Not Compensated

E45-319 · Calibration and Adjustment · Gas Mass Flow Meter · ★

Problem Description

Atmospheric pressure was not compensated; after calibration the volumetric flow reads 3%FS off

Root Cause

[Mechanism] Gas flow calibration must use absolute pressure as the reference; not compensating the local atmospheric pressure makes the working-pressure reference wrong and biases the volumetric conversion.
[Calculation] The local atmospheric pressure of 101.3kPa was not compensated, making the working pressure off by 3% and the volumetric flow off by 3%FS.
[Design Omission] No atmospheric-pressure compensation was taken and no absolute-pressure conversion was made.

Corrective Solution

① Pressure compensation: compensate with the local atmospheric pressure → true reference
② Absolute pressure: convert to absolute pressure → correct value
③ Verification: recalibrate at a different air pressure → confirm consistency
④ Out-of-tolerance alarm: warn when the deviation >1% → early warning
⑤ Records: keep the air pressure on file → traceability
⑥ Periodic calibration: calibrate every quarter → prevent drift
⑦ Recalibration: recalibrate when the altitude changes → correct the bias
⑧ Write the air-pressure requirement into the procedure → control at the source

Checklist

Reference

JJG 1030-2007 Verification Regulation of Ultrasonic Flowmeters