E45-318 · Calibration and Adjustment · Electromagnetic Flow Meter · ★★
Calibrated with water but measuring slurry, the reading is off by 8%; the coefficients do not match
[Mechanism] The calibration coefficient of an electromagnetic flow meter depends on the conductivity of the medium; after calibrating with water (high conductivity) and then measuring a low-conductivity slurry, the signal characteristics change and the coefficient becomes inaccurate.
[Calculation] Water calibrates at 500μS/cm and the slurry is 50μS/cm; the 8% coefficient difference biases the reading.
[Design Omission] Calibration was not done with the actual medium and no conductivity compensation was applied.
① Medium calibration: calibrate with the actual medium → the coefficient fits
② Conductivity compensation: add a conductivity compensation term → cancel the difference
③ Verification: compare against actual field flow → confirm consistency
④ Out-of-tolerance alarm: warn when the deviation >2% → early warning
⑤ Records: keep the medium on file → traceability
⑥ Periodic calibration: calibrate every quarter → prevent drift
⑦ Recalibration: recalibrate when the medium changes → correct the bias
⑧ Write the medium requirement into the procedure → control at the source
JJG 1033-2007 Verification Regulation of Electromagnetic Flowmeters