E45-317 · Calibration and Adjustment · Electromagnetic Flow Meter · ★★
The weighing-method calibration did not tare the container; after calibration the flow reads 5% off
[Mechanism] The weighing-method calibration did not deduct the container tare or the accumulation time was inaccurate, making the calibration reference mass wrong and biasing the derived flow coefficient.
[Calculation] The 2kg container tare was not deducted, so a nominal 100kg was actually 98kg, biasing the flow by 2%; the timing error adds another 3%.
[Design Omission] No taring, no time calibration, and no verification.
① Taring: tare the container before calibration → true net mass
② Time calibration: calibrate the accumulation time → no bias
③ Re-weighing: re-weigh after calibration to verify → confirm consistency
④ Second method: recalibrate by the volumetric method → cross-check
⑤ Out-of-tolerance alarm: warn when the deviation >2% → early warning
⑥ Records: keep the tare on file → traceability
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the weighing-method requirement into the procedure → control at the source
JJG 1033-2007 Verification Regulation of Electromagnetic Flowmeters