Leakage in Vacuum Equipment

E46-245 · Bonding and Sealing · Vacuum Chamber · ★★★

Problem Description

An ordinary sealant was used on the joint face of the vacuum chamber; after evacuation the micro-pores in the adhesive layer leaked, the vacuum level never met the requirement, and the pressure dropped rapidly during the hold test.

Root Cause

[Mechanism] Ordinary sealant is gas-permeable and contains micro-pores; under vacuum, gas permeates through it or leaks via the pores, so the vacuum level is not reached.
[Calculation] The adhesive's permeability exceeds the limit, making the leak rate several times the value allowed for the equipment.
[Design gap] No vacuum sealant grease or metal gasket was used on the vacuum equipment.

Corrective Solution

① Use vacuum seals: use vacuum sealant grease or metal gaskets at vacuum joints
② Control leakage: choose low-permeability sealing materials
③ Metal gaskets: use metal gaskets such as oxygen-free copper for high vacuum
④ Clean faces: keep joint faces highly finished and free of scratches
⑤ Even tightening: tighten the bolts evenly
⑥ Helium test: confirm with helium mass-spectrometer leak detection
⑦ Seal verification: confirm the vacuum level meets the specification by a vacuum test
⑧ Standardize process: write the vacuum sealing materials into the design

Checklist

Reference

GB/T 6070 Vacuum technology - Flange dimensions

Related model download:Understood the principle and ready to try?Go to MOULDL to download related 3D models and CAD drawings for free »