Excessive CT Gantry Bearing Hole Position Blurs the Image

E43-245 · Geometrical Tolerance Errors · Medical CT Gantry Bearing Bore · ★★★

Problem Symptom

Bearing hole position Φ0.25mm (allowable Φ0.08) made the rotation uneven and blurred the image

Root Cause

[Mechanism] Zero point drift on the machining center shifted the hole positions.
[Calculation] A position tolerance of Φ0.25 shifted the holes by 0.25 against the 0.08 allowable, so the radial runout during rotation became large.
[Design Omission] No on-line tool setting was performed and the position tolerance was not inspected; this design flaw blurred the image.

Correct Solution

1) Use on-line tool setting on the machining center;
2) Inspect the position tolerance ≤Φ0.08;
3) Unify the datum;
4) Measure the first piece on a CMM;
5) Select P4 bearings;
6) Inspect the radial runout during rotation;
7) Verify the image resolution;
8) Bring position tolerance into the mandatory inspection list.

Checklist

References

GB/T 1182-2018 Geometrical Tolerances

Related 3D models: Ready to build it? Download free 3D models & CAD drawings at MOULDL »
MOULDL · Mechanical Engineering Knowledge Base · free self-study reference for machine designers, 3D CAD models at MOULDL