E44-469 · Filter Clogging and Contamination · Sterilizing-Grade Filter 0.22μm · ★★★
Excessive drying made the hydrophilic cartridge hydrophobic; the bubble point dropped from 3.5bar to 1.8bar<2.5, integrity failed, and sterility assurance was lost
[Mechanism] Excessive drying temperature made the membrane hydrophobic, swelled its pore size, and lowered the bubble point.
[Calculation] Bubble point P=2γcosθ/d; a change in θ drops P from 3.5 to 1.8<2.5, breaking the barrier.
[Design Omission] No control of drying temperature and no per-batch bubble-point test, so integrity failed and bacteria bypassed.
1) Limit drying temperature to ≤60°C to preserve hydrophilicity; since overheating changed θ and gave P=1.8<2.5, temperature control preserves P=3.5;
2) Perform a per-batch bubble-point test at ≥2.5bar; since failure lets bacteria through, batch testing catches it;
3) Store cartridges wet; since drying makes them hydrophobic, wet storage keeps them hydrophilic;
4) Switch to hydrophilic dry-tolerant PVDF; since the original material was not dry-tolerant, the new material is;
5) Use saturated steam for sterilization to avoid dry baking; since dry baking damages the membrane, wet sterilization protects it;
6) Test integrity both before and after installation; since installation can also damage the membrane, testing catches it;
7) Keep batch records for traceability; since pharmaceutical products require traceability, records are maintained;
8) Write the bubble-point test into the aseptic procedure; since the lack of testing was the root cause of failure, the aseptic procedure provides the check.
GB/T 13554-2020 High-efficiency air filters