MOULDLKnowledge Base3D Model DownloadsMembership
中文版Model Store »
HomeModels
Table of contents

Knowledge Base Home

📐 Standard Data

🧮 Calculators

📖 Selection Guides

🏭 Material Grades

❓ Engineer FAQ

🗂 Quick-Reference Categories

❌ Error Case Studies

📚 Technical Articles

Knowledge Base › Electrical Sensor Error Case › Overly Large Weighing Range Leaves Insufficient Resolution and Causes Batching Deviation

Overly Large Weighing Range Leaves Insufficient Resolution and Causes Batching Deviation

E45-002 · Sensor Selection and Range · Load cell HBM C16 0-500kg · ★★

Problem Description

A load cell with a 0~500 kg range was used to weigh an actual 5 kg small ingredient; the resolution 0.1%FS = 0.5 kg is more than 10% of the ingredient weight, so batching deviated ±10% beyond tolerance and product composition was unstable

Root Cause

[Mechanism] The sensor resolution is 0.1%FS = 0.5 kg; the 5 kg ingredient being weighed is only 1% of the range, and the quantization step is more than 10% of the ingredient weight, making the reading jump.
[Calculation] The noise fraction rises at small signals, further degrading effective resolution to a repeatability error of ±0.8 kg.
[Design Omission] The range was not selected for 10~100% of the weighed amount and no staged weighing was used; the range is too large, leaving insufficient resolution.

Corrective Solution

① Range matching: 5 kg ingredient being weighed → select a 0~10 kg range (full scale ≥ ingredient weight × 1.2); the ingredient weight falls near 50% of range, giving a resolution of 0.01 kg and a deviation ≤1%
② Staged weighing: use a 500 kg cell for the bulk and a 10 kg cell for the small ingredient → each is in its optimal range segment and neither degrades the other's resolution
③ Resolution selection: choose a model with resolution ≤0.02%FS → the original 0.1%FS gives a 0.5 kg step at 5 kg, which is too coarse; this cuts the quantization error to an acceptable level
④ Digital filtering: zero tracking + averaging over 16 samples → suppress the high noise fraction at small signals and improve repeatability from ±0.8 kg
⑤ Eccentric-load calibration: full-scale three-point calibration + four-corner eccentric-load test → four-corner error ≤0.05%FS so small-ingredient readings do not skew
⑥ Tolerance alarm: stop when the batching deviation >2% → promptly reject the ±10% out-of-tolerance batch so it does not flow to the next process
⑦ Weight verification: test-weigh with a 5 kg standard weight, reading 5.0±0.05 kg → confirm on site that the resolution is adequate instead of relying on design estimates
⑧ Lock in selection: write the range matching into the hopper-scale datasheet as a mandatory item → eliminate mismatch-induced insufficient resolution at the source

Checklist

Reference

GB/T 7723-2008 Stationary electronic weighing instruments

Related model download:Understood the principle and ready to try?Go to MOULDL to download related 3D models and CAD drawings for free »
MOULDL · Mechanical Engineering Knowledge Base · free self-study reference for machine designers, 3D CAD models at MOULDL