E45-022 · Installation Position and Orientation · Vortex Flowmeter on a DN80 Compressed-Air Line · ★
The flow-arrow marking was reversed during installation, so it reads −5% or zero and gas usage cannot be measured
[Mechanism] A vortex flowmeter needs flow in the correct direction to produce a stable Karman vortex street; reversed, the vortex shedding is disordered and the processor judges it invalid.
[Calculation] Reversed, the Strouhal number St breaks down and the shedding frequency f = St·v/d has no solution, so the output clamps to zero or offsets negative.
[Design Omission] The flow arrow was not checked, no flow test was run, and the medium direction was not marked.
① Reinstall with the flow: reinstall along the flow as the body arrow shows (was reversed) → reversed, the vortex street was disordered with no stable frequency; with the flow it produces a measurable frequency
② Direction marking: mark the medium flow arrow on the pipe section to prevent reversal → easy to mix up; make it explicit
③ Flow test: run a flow test before installation to see a positive reading → design often skips it; confirm by measurement
④ Bidirectional meter: choose a bidirectional vortex meter (symmetric bluff body) as a spare → field flow may be reversed; support both directions
⑤ Straight run: keep an upstream straight run ≥15D to steady the flow field → distortion upsets the field and makes the reading jump; steady the reading
⑥ Direction check: verify the direction and reading each quarter → re-piping can reverse it; correct it early
⑦ Downstream comparison: compare with the downstream meter (difference <2%) to verify → a zero reading is hidden and hard to spot; cross-check with calibration
⑧ Make it stick: put the flow-direction check into the installation card → eliminates reversed-install zero readings at the source
GB/T 25922 Vortex flowmeters