E45-077 · Signal Interference and EMC · Platinum Resistance Thermometer · ★
After a nearby large motor starts, the temperature display drifts ±3℃ (it was stable within ±0.2℃) and recovers as soon as the motor stops; the design limit is ≤±0.5℃
[Mechanism] The 50Hz mains-frequency magnetic field induces an EMF in the sensor signal loop, forming common-mode interference that appears as temperature drift after unbalanced amplification.
[Calculation] With signal-loop area A≈5cm² and B≈0.5mT, the induced e=2πf·B·A≈78μV; through high gain and unbalanced amplification it rises to ±3℃.
[Design Omission] Three-wire compensation was not used, the conductors were not twisted, and the cable was not kept away from high current.
① Twisted conductors: change the signal cable to twisted pair (lay length ≤25mm) → the loop area is halved and induction drops 75%
② Three-wire connection: switch the platinum resistor to three-wire compensation of the lead resistance → eliminates the common-mode drift
③ Keep away from high current: keep the sensor cable ≥200mm from power lines → the magnetic field attenuates
④ Shielding: add a shield grounded at one end → blocks electric fields
⑤ Mains notch: add a 50Hz notch filter (depth ≥40dB) → drift is pressed down to ±0.3℃
⑥ Balanced amplification: use a front-end instrumentation amplifier with common-mode rejection ratio ≥100dB → common mode does not convert into differential mode
⑦ Verification by test: compare against a standard source with the motor running (difference ≤0.5℃) → confirm
⑧ Institutionalize: write the temperature-cable routing into the card → gatekeeping at the source
GB/T 17626.6-2017 Immunity to conducted disturbances induced by radio-frequency fields