E45-091 · Signal Interference and EMC · pH Electrode · ★★
After a nearby high-voltage cable is energized the pH reading drifts by 0.5 pH (previously stable within ±0.05 pH)
[Mechanism] Capacitive coupling between the high-voltage cable and the high-impedance pH electrode line couples electric-field interference; through the high-impedance electrode it introduces a spurious current that drifts the reading.
[Calculation] With coupling capacitance C≈5pF and interference voltage U=1kV, i=C·du/dt≈a few nA; across the high pH impedance (10^9Ω) this gives a ΔU equivalent to a 0.5 pH drift.
[Design Omission] The pH line was not shielded, not kept away from the high voltage, not filtered, and not protected.
① Shielding and twisted pair: change the pH line to shielded twisted pair with the shield grounded → blocks electric-field coupling
② Keep away from high voltage: maintain ≥0.5m from the high-voltage cable → reduces capacitive coupling
③ High-impedance protection: add a guard ring at the electrode input → suppresses leakage current
④ Low-pass filtering: add a low-pass filter (cutoff 1Hz) → filters high frequencies
⑤ Independent floating ground: float the measuring end and add isolation → reduces common mode
⑥ Shielded box: use a metal-shielded electrode junction box → blocks radiation
⑦ Verification by test: compare pH with the high voltage energized (drift ≤0.05 pH) → confirm
⑧ Institutionalize: write the high-impedance signal routing into the card → gatekeeping
GB/T 17626.6-2017 Immunity to conducted disturbances induced by radio-frequency fields