E45-108 · Signal Interference and EMC · Electromagnetic Flowmeter · ★★
Near an induction furnace the electromagnetic flowmeter reading jumps erratically by ±10%FS
[Mechanism] The strong magnetic field of the induction furnace (several kHz) induces an EMF in the flowmeter electrodes and leads that swamps the flow signal.
[Calculation] With induction field B≈5mT at 2kHz, the EMF induced in the electrode loop area A, e=2πfBA, makes the reading jump ±10%FS.
[Design Omission] The flowmeter cable was not shielded, not kept away, not filtered and not magnetically shielded.
① Magnetic shielding: fit the electrode leads with an iron tube / permalloy sleeve → attenuates the external field by 20dB
② Move away from the furnace: keep at least 2m from the induction furnace → reduces B
③ Shielded twisted pair: run the signal cable as shielded twisted pair → suppresses coupling
④ Low-pass filtering: add a low-pass filter (10Hz cutoff) → filters out the induction frequency
⑤ Isolation: isolate the signal → cuts common mode
⑥ Grounding: ground the shield at one point → avoids ground loops
⑦ Verify by measurement: readings with the furnace running (fluctuation ≤1%) → confirmed
⑧ Institutionalize: write keeping the flowmeter away from induction sources into the standard → gate at the source
GB/T 17626.3-2016 Radiated radio-frequency electromagnetic field immunity test