E45-187 · Grounding and Shielding · Load Cell · ★★
The weighing reading drifts slowly by ±0.5% with no pattern; the sensor signal ground is connected to no ground at all
[Mechanism] With the signal ground floating, the sensitive circuit has no stable reference potential; stray parameters influenced by the environment (humidity, nearby live parts) make the reference float and add to the bridge output; floating also tends to accumulate static charge.
[Calculation] Floating capacitance to ground C≈10pF and environmental charge Q give ΔU=Q/C; a static charge of 1nC → ΔU on the order of 100V, which through the high-impedance bridge works out as reading drift.
[Design Omission] The weighing module does not establish a reference by bonding the signal ground to protective earth at a single point.
① Single-point PE bond: bond the signal ground to protective earth at one point (through 1MΩ or a capacitor) → establishes a reference without circulating current
② Bridge shielding: bond the sensor cable shield to the signal ground → resists coupling
③ Equipotential: bond the instrument ground to the cabinet PE at one point → unified reference
④ Bleed resistor: add 1MΩ from the signal ground to PE to bleed static charge → no accumulation
⑤ Filtering: apply a long-time-constant software filter (τ=2s) → suppresses slow drift
⑥ Reference test: measure the signal-ground-to-PE resistance ≈1MΩ with a stable potential → verify
⑦ Drift ratio: long-term no-load drift <0.1% → verify
⑧ Write the grounding reference into the instrument manual → gate at the source
GB/T 7724-2008 Electronic weighing instruments