MOULDLKnowledge Base3D Model DownloadsMembership
中文版Model Store »
HomeModels
Table of contents

Knowledge Base Home

📐 Standard Data

🧮 Calculators

📖 Selection Guides

🏭 Material Grades

❓ Engineer FAQ

🗂 Quick-Reference Categories

❌ Error Case Studies

📚 Technical Articles

Knowledge Base › Electrical Sensor Error Case › Crosstalk Between Power Ground and Signal Ground

Crosstalk Between Power Ground and Signal Ground

E45-210 · Grounding and Shielding · Integrated Acquisition Instrument · ★★

Problem Description

The instrument's power ground and signal ground were shared, so supply ripple coupled into the measurement and made the reading jitter by ±0.3%

Root Cause

[Mechanism] The ground of a switching supply carries high-frequency noise between primary and secondary; if the power ground and signal ground share a section directly, that noise couples to the signal-ground reference through the common impedance and adds to the measured value.
[Calculation] With supply ripple ΔU=50mV, a shared-section impedance of 0.02Ω and a noise current of 2A, the coupling to the signal-ground reference makes the reading jitter by ±0.3%.
[Design Omission] The power ground and signal ground were neither separated nor isolated.

Corrective Solution

① Ground separation: route the power ground and signal ground separately → no shared section
② Isolated supply: use an isolated DC-DC converter (1500V) → decouples the two grounds
③ Single-point tie: connect the two grounds at one point through a Y capacitor → passes RF, blocks mains frequency
④ π filter: add a π-type filter at the supply input (10Ω + 10μF) → suppresses ripple
⑤ Separate planes: lay out the power and signal grounds on separate PCB layers → less coupling
⑥ Jitter test: measure reading jitter <0.1% → confirmed
⑦ Ripple test: signal-ground ripple <10mV → confirmed
⑧ Write the ground-separation requirement into the instrument design → control at the source

Checklist

Reference

GB/T 9254-2008 Radio disturbance of information technology equipment

Related model download:Understood the principle and ready to try?Go to MOULDL to download related 3D models and CAD drawings for free »
MOULDL · Mechanical Engineering Knowledge Base · free self-study reference for machine designers, 3D CAD models at MOULDL