E45-277 · Calibration · Level Transmitter · ★
The reference pressure used for calibration was below the lower range limit; the zero calibration became inaccurate and the whole range shifted by 2%FS
[Mechanism] The reference pressure point was chosen at or beyond the edge of the sensor's range, so the reference point does not represent the working segment; the calibration coefficient is derived from a wrong basis and the whole measurement shifts.
[Calculation] With a range of 0~10kPa the calibration point was taken at -2kPa (outside the range); the slope derived this way is off by 2% and the whole 0~10kPa working segment is biased.
[Design Omission] It was not checked that calibration points must lie within the range, and no representative point of the working segment was chosen.
① In-range points: calibration points must lie within 0~100%FS → correct basis
② Representative points: take 25/50/75% of the working segment → match reality
③ Range check: verify the range match before calibration → stay within limits
④ Recheck: compare the full-scale point after calibration → verify the slope
⑤ Out-of-range alarm: alarm when a point is out of range → prevent wrong calibration
⑥ Records: keep the calibration points on file → traceability
⑦ Re-calibration: re-calibrate on any anomaly → correct the deviation
⑧ Write the calibration rules as constraints → control at the source
GB/T 17614.1-2015 Industrial process control systems - Transmitters - Part 1: Methods of evaluating the performance