MOULDLKnowledge Base3D Model DownloadsMembership
中文版Model Store »
HomeModels
Table of contents

Knowledge Base Home

📐 Standard Data

🧮 Calculators

📖 Selection Guides

🏭 Material Grades

❓ Engineer FAQ

🗂 Quick-Reference Categories

❌ Error Case Studies

📚 Technical Articles

Knowledge Base › Electrical Sensor Error Case › Zero Drift Without Temperature Compensation

Zero Drift Without Temperature Compensation

E45-278 · Calibration · Thermocouple · ★

Problem Description

The cold junction was not compensated and no temperature calibration was done; a 15℃ ambient change offset the reading by 0.6mV, about 15℃

Root Cause

[Mechanism] A thermocouple's EMF depends on the cold-junction temperature; without compensation the cold junction drifts with the ambient, and the moving reference shifts the whole range of readings.
[Calculation] A type-K thermocouple gives 0.04mV/℃; a 15℃ temperature change introduces a 0.6mV error, about a 15℃ reading offset.
[Design Omission] No cold-junction compensation was fitted and the calibration contained no temperature-compensation term.

Corrective Solution

① Cold-junction compensation: fit a cold-junction compensator → stable reference
② Temperature-compensated calibration: include a temperature correction term → drift removed
③ Isothermal calibration: hold the cold junction at 0℃ during calibration → accurate reference
④ Real-time compensation: measure the cold-junction temperature and correct dynamically → accurate with temperature
⑤ Re-calibration: re-calibrate after a temperature change → tracking
⑥ Out-of-tolerance alarm: alarm when the error >1℃ → early warning
⑦ Periodic calibration: calibrate every quarter → prevent drift
⑧ Write the compensation into the procedure → control at the source

Checklist

Reference

GB/T 16839.1-2018 Thermocouples - Part 1: EMF specifications and tolerances

Related model download:Understood the principle and ready to try?Go to MOULDL to download related 3D models and CAD drawings for free »
MOULDL · Mechanical Engineering Knowledge Base · free self-study reference for machine designers, 3D CAD models at MOULDL