E45-279 · Calibration · Pressure Sensor · ★★
Calibrated with an out-of-tolerance reference gauge; the sensor error actually grew to 2%FS and traceability was broken
[Mechanism] The reference standard being out of tolerance makes the calibration basis wrong; the sensor is calibrated to a wrong reference point, the error is amplified rather than removed, and the traceability chain breaks.
[Calculation] The gauge error is 1%FS and the sensor was originally 0.5%FS; calibrated against the wrong reference it ends up with a 1% error, for a total of 2%.
[Design Omission] The reference's validity and certificate were not checked and the reference was not self-checked.
① Validity check: check that the reference certificate is current before calibration → true basis
② Self-check: self-check the standard periodically → keep it accurate
③ Class matching: the standard's accuracy 3 times better than the unit → true value
④ Certificates: keep the certificate numbers → traceability
⑤ Recheck: re-calibrate with another standard → confirm agreement
⑥ Out-of-tolerance alarm: forbid calibration when the standard is out of tolerance → prevent wrong calibration
⑦ Management: keep a standards log → manage the intervals
⑧ Write traceability into the rules → control at the source
JJF 1001-2011 General Terms in Metrology and Their Definitions