Knowledge Base › Electrical Sensor Error Case › Wrong Range Calibration Coefficient

Wrong Range Calibration Coefficient

E45-281 · Calibration · Vortex Flowmeter · ★

Problem Description

The full-scale coefficient was entered wrong during calibration; the full-scale reading ran high by 8%, overstating the measurement

Root Cause

[Mechanism] A manual entry error in the range coefficient after calibration shifts the linear slope from its true value; the whole range is offset proportionally and the output no longer matches the actual flow.
[Calculation] The true full scale is 100m³/h but 108m³/h was entered; the coefficient is off by 8% and the whole range is scaled by 108/100.
[Design Omission] The coefficient was not double-checked and no actual-flow verification was done.

Corrective Solution

① Two-person check: have two people check the coefficient entry → no entry error
② Actual-flow verification: verify with a reference meter on actual flow after calibration → verify the slope
③ Range lock: changing the coefficient requires authorization → prevent error
④ Comparison: compare against the production meter after calibration → confirm agreement
⑤ Recheck: recalculate at different flow points → accurate over the whole range
⑥ Error alarm: alarm when the deviation >2% → early warning
⑦ Records: keep the coefficient on file → traceability
⑧ Write the entry rules as constraints → control at the source

Checklist

Reference

JJG 1029-2007 Verification Regulation of Vortex Flowmeters